Assessing The Efficiency Of Home Glucose Monitors In Comparison To Standard Laboratory Testis

Authors

  • Ahmed Ali Mohalhel Department of Medical Sciences, Higher Institute of Science and Technology, Zliten Author
  • Alsideeq Abduasalam Aboujanah Department of Medical Sciences, Higher Institute of Medical Technology,Emsalata Author

DOI:

https://doi.org/10.65405/wyjrkv46

Keywords:

Diabetes / Blood glucose / Glucose / Home glucose meter.

Abstract

Background: Glucose is the main source of energy in the body, and its regulation is controlled by a hormonal balance between insulin and glucagon. This balance ensures a stable blood glucose level to support the body's various functions.

Objective: The purpose of this study is to determine the accuracy of home blood glucose meters in measuring blood sugar levels compared to the results of laboratory devices.

Materials and Methods: This study included 115 samples collected between November 2024 and January 2025 from the Diabetes and Endocrinology Treatment Center in Msallata. The ages of the sample donors ranged from 5 to 93 years. The Accu-Chek Active home glucose meter and the laboratory of the same center were used to measure all the samples.

Results: The results of this study indicated a slight increase in the laboratory test results compared to the home glucose meter readings. The average laboratory test results were 212.39 mg/dL, while the average readings from the home glucose meter were 192.28 mg/dL, with a difference of 20.11 mg/dL between the two measurement methods.

Conclusion: The findings of this study suggest that while there are slight differences between home glucose meters and laboratory tests, home glucose meters can be effectively relied upon for monitoring blood sugar levels in diabetic patients, provided that proper usage techniques are followed and the manufacturer's instructions are adhered to.

Downloads

Download data is not yet available.

References

1. Al-Ansari, M., & Al-Malki, A. S. (2022). Accuracy and clinical agreement of capillary point-of-care blood glucose meters compared with laboratory venous plasma glucose testing in diabetic patients. Journal of Clinical & Diagnostic Research, 16(8), BC01–BC06.

2. Badireddy, M., & Deepthi, R. (2023). Hyperglycemia. StatPearls Publishing.

3. Gifford-Jorgensen, R. A., Borchert, J., Hassanein, R., Tilzer, L., Eaks, G. A., & Moore, W. V. (1986). Comparison of five glucose meters for self-monitoring of blood glucose by diabetic patients. Diabetes Care, 9(1), 70–76.

4. Gillery, P., et al. (2015). Glucose meters - Fit for clinical purpose. Clinical Chemistry and Laboratory Medicine, 53(10), 1–10.

5. Halder, P., et al. (2024). Reliability and Accuracy of Bedside Capillary Blood Glucose Measurement by Glucometers Compared to Venous Blood Glucose in Critically Ill Patients. Journal of Clinical and Diagnostic Research.

6. International Diabetes Federation. (2017). IDF Diabetes Atlas (8th ed.). International Diabetes Federation.

7. Mathew, T. K., Zubair, M., & Tadi, P. (2023). Blood Glucose Monitoring. StatPearls Publishing.

8. Tirimacco, R., Koumantakis, G., Erasmus, R., Mosca, A., Sandberg, S., Watson, I. D., Goldsmith, B., & Gillery, P. (2015). Glucose meters - Fit for clinical purpose. Clinical Chemistry and Laboratory Medicine, 51(5), 1–10.

9. World Health Organization. (2020). Classification of diabetes mellitus. World Health Organization.

Downloads

Published

2026-11-28

How to Cite

Assessing The Efficiency Of Home Glucose Monitors In Comparison To Standard Laboratory Testis . (2026). Al-Farooq Journal of Sciences, 2(5), 278-284. https://doi.org/10.65405/wyjrkv46